Safety and Regulatory Information
This product complies with EN/IEC 61326-2-1, for sensitive test and measurement equipment:
When subjected to transient radiated and/or conducted electromagnetic phenomena, the
product may have temporary loss of function or performance which is self-recovering. Recov-
ery may take longer than 10 seconds.
When subjected to continuously present electromagnetic phenomena, some degradation of
performance may occur.
Unless otherwise noted in the specifications, this instrument or system is intended for indoor
use in an installation category II, pollution degree 2 environment per IEC 61010-1 and 664
respectively. It is designed to operate at a maximum relative humidity of 5% to 80% at 40 °C or
less (non-condensing). This instrument or system is designed to operate at altitudes up to 3000
meters, and at temperatures between 0 and 55 °C.
This instrument is used with mains power of 100 - 240 V, 50/60 Hz or 100 - 120 V, 400 Hz. The
maximum power usage is 25 VA. Mains supply voltage fluctuations are not to exceed ±10% of
the nominal supply voltage. Ensure you are using the correct mains rated cord based one the
instrument's rating and your country's electrical codes.
Product Regulatory and Compliance
The Truevolt Series DMMs comply with safety and EMC requirements.
Refer to Declaration of Conformity at
Measurement Limits
The Truevolt Series DMMs provide protection circuitry to prevent damage to the instrument and to pro-
tect against the danger of electric shock, provided the Measurement Limits are not exceeded. To ensure
safe operation of the instrument, do not exceed the Measurement Limits shown on the front and rear
panel, and defined as follows:
34460A Front Panel
Terminal Area
16
http://www.keysight.com/go/conformity
Keysight Truevolt Series Operating and Service Guide
for the latest revision.