CALibration:DATE?
Returns the date of the last calibration in the format yyyy,mm,dd.
Parameter Typical Return
(none)
2014,4,26
Return the calibration date:
CAL:DATE?
You can perform this query regardless of whether the instrument is secured.
l
The date is based on the instrument's real-time clock. Set the date for the instrument's real-time
l
clock with
SYSTem:DATE
This setting is non-volatile; it will not be changed by power cycling or
l
See Also
SYSTem:DATE
SYSTem:TIME
CALibration:SECure:CODE <new_code>
Sets the security code (calibration password) to prevent unauthorized calibrations.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
Unquoted string up to 12 characters
Must start with letter (A-Z)
May contain letters, numbers (0-9) and underscores
Set new security code (calibration password):
CAL:SEC:STAT OFF,OLD_CAL_CODE
CAL:SEC:CODE TST_DUT165
CAL:SEC:STAT ON
To change the code: unsecure calibration memory with the old code, then set the new code.
l
If you forget your security code, see
l
When shipped from the factory, the instrument is secured, with the security code (calibration pass-
l
word) set to AT3446XA.
This setting is non-volatile; it will not be changed by power cycling or
l
Keysight Truevolt Series Operating and Service Guide
and SYSTem:TIME.
(none)
Security Code
SCPI Programming Reference
*RST
Typical Return
Override.
*RST
or SYSTem:PRESet.