Features and Functions
Capacitance
This section describes how to configure capacitance measurements from the front panel.
Step 1: Configure the test leads as shown.
Step 2: Press
Step 3: To null–out the test lead capacitance:
Disconnect the + and - test leads probe end from the test circuit, and leave open.
l
Press Null. The DMM will now subtract this null value from capacitance measurements.
l
Step 4: Press Range to select a range for the measurement. You can also use the [+], [-], and [Range]
keys on the front panel to select the range. (Auto (autorange) automatically selects the range for the
measurement based on the input. Autoranging is convenient, but it results in slower measurements than
using a manual range. Autoranging goes down a range at less than 10% of range and up a range at
greater than 120% of range. For capacitance measurements only, when autorange is off, the instrument
does not report an overload for readings greater than 120% of range. Overload only occurs when the
algorithm times out because the applied capacitance is too large for the algorithm to measure. If you
apply a DC voltage or a short to the input terminals in capacitance measurement mode, the instrument
reports an overload.
76
on the front panel.
Keysight Truevolt Series Operating and Service Guide