Clearing Reading Memory
The following actions clear reading memory:
Changing the measurement function
l
Pressing any Clear Readings softkey
l
Transitioning in or out of Probe Hold
l
Changing temperature units
l
Changing any dB/dBm parameters
l
Changing any histogram binning parameter
l
Changing the temperature probe or R
l
Recalling a stored state
l
Calibrating the instrument
l
Switching between 3 A and 10 A inputs
l
Changing the position of the Front/Rear switch
l
Transferring from Remote to Local mode
l
Turning Null on or off or changing the Null value
l
These actions do not clear reading memory:
Changing measurement parameters, such as range and aperture.
l
Turning limits on or off, or adjusting limit values
l
Pressing [Run/Stop] in front panel auto trigger mode
l
Changing the trend chart Recent/All softkey (34461A/65A/70A only)
l
Changing samples per trigger or trigger delay
l
Changing display modes
l
Changing the VM Comp output polarity
l
Changing the digit mask
l
Changing the histogram, bar meter, or trend chart scale
l
Changing user preferences
l
Executing self-test