AC Zero Calibration
This calibration procedure applies to all models (34460A/61A/65A/70A).
Review
Test Considerations
Configuration: AC Voltage
1. Set the calibrator to 0.00 Ohm.
2. Enter a calibration value of +0. (see
sage indicates success; if the display shows Calibration Step Failed, check the input value, range,
function, and entered calibration value and repeat the calibration step.
3. Store the new calibration constants (see
This calibration takes approximately 10 seconds to complete.
Keysight Truevolt Series Operating and Service Guide
before performing any calibration procedure.
Entering Calibration
Storing the Calibration
Calibration Adjustment Procedures
Values). A Calibration Step Succeeded mes-
Constants).
Next Procedure