Guidance and manufacturer's declaration - electromagnetic immunity
The InZone
Detachment System is intended for use in the electromagnetic environment specified below. The customer or the user of the InZone
®
Detachment System should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
a
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the InZone Detachment System is used
exceeds the applicable RF compliance level above, the InZone Detachment System should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or relocating the InZone Detachment System.
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of
the InZone Detachment System, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
recommended separation distance
Not Applicable
Not Applicable
3 V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range.