Guidance and manufacturer's declaration - electromagnetic immunity
The InZone
Detachment System is intended for use in the electromagnetic environment specified below. The customer or the user of the InZone
®
Detachment System should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
U
NOTE:
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
Compliance level
± 2, 4, 6 kV contact
± 2, 4, 6 kV contact
± 2, 4, 8 kV air
± 2, 4, 8 kV air
± 2 kV for power supply lines
Not applicable
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
Not applicable
± 2 kV line(s) to earth
U
<5 %
T
U
(>95 % dip in
)
T
for 0.5 cycle
U
40 %
T
U
(60 % dip in
)
T
for 5 cycles
Not applicable
U
70 %
T
U
(30 % dip in
)
T
for 25 cycles
U
<5 %
T
U
(>95 % dip in
)
T
for 5 s
3 A/m
3 A/m
10
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.