B) Attachment Wraps
Body Weight
1 kg ~ 3 kg
3 kg ~ 10 kg
10 kg ~ 30 kg
10 kg ~ 50 kg
> 30 kg
C) Attaching the adhesive squares to the sensor (not required for CleanShield Multisite Wrap)
NOTE: For improved adherence of the adhesive squares to the sensor wipe the sensor pads with 70% isopropyl alcohol and
allow to dry prior to attaching the adhesive squares.
1. Refer to Fig. 1a. Remove the adhesive squares from the backing
2. Refer to Fig. 1b. Attach one square to each window of the sensor pads (emitter and detector). Avoid touching the sticky
side prior to applying to the sensor pads.
3. Do not remove the release liner until ready to apply the sensor to the site.
CAUTION: Do not use adhesive squares on fragile skin.
D) Attaching the Sensor to the Patient
Refer to Fig. 2. The foam wrap can be trimmed to the shorter length for smaller site applications such as a child's finger or
toe or a pre-term infant's foot or hand.
1. Remove the backing from the adhesive wrap, leaving the backing on the tab end (Standard Wrap, CleanShield Wrap and
Standard Petite Wrap only).
2. Refer to Fig. 3a. Push the "button" on the emitter sensor pad (cable is marked with a red indicator) through the hole on
the left Fig. 3b. and the remaining button on the detector sensor pad through the hole on the right Fig 3c.
NOTE: Use the center hole of the Petite Wrap for patients 10 – 30 kg. Refer to Fig 3d.
3. Refer to Fig. 4. Fold the upper portion of the attachment tape down over the sensor pads and remove the release liner
from the folded over section of the adhesive wrap prior to applying the sensor to the site (CleanShield Wrap only).
NEONATES (1-3 kg) and INFANTS (3-10 kg), Standard Wrap / CleanShield Wrap / Foam Wrap
1. Direct the sensor cable toward the patient. Orient the YI on the outer aspect of the foot Fig. 5a. or hand Fig. 5b. with the
center of the wrap Fig. 5c. under the 5th digit and the detector window on the fleshy portion and the emitter window
(cable is marked with a red indicator) directly opposite.
2. Secure the small tape end and remove the remaining backing (Standard Wrap, CleanShield Wrap and Standard Petite
Wrap only).
3. Wrap the tape loosely enough to avoid restricting circulation around the site and to maintain proper alignment of the
detector and emitter windows; foot Fig 6a., hand Fig. 6b.
4. If using the Foam Wrap, secure the Foam Wrap Fig. 6c. using the hook and loop tab.
NOTE: When placing the YI on the hand, it is recommended that the EMITTER (cable is marked with a red indicator) pad be
placed on the palm of the hand, under the 4th or 5th finger.
ADULTS (> 30 kg) and PEDIATRICS (10-50 kg), Standard Wrap / CleanShield Wrap / Petite Wrap / Foam Wrap
1. Refer to Fig. 7a. Direct the sensor cable toward the patient. Orient the YI on the selected site so the detector window is
on the fleshy portion and the emitter window (cable is marked with a red indicator) directly opposite on the nail bed.
2. Refer to Fig. 7b. Secure the small tape end and remove the remaining backing (Standard Wrap, CleanShield Wrap and
Standard Petite Wrap only).
3. Refer to Fig. 7c. Wrap the adhesive or foam wrap around the finger loosely enough to avoid restricting circulation
around the site and to maintain proper alignment of the detector and emitter windows.
NOTE: When placing the YI on a finger or toe, the EMITTER (cable is marked with a red indicator) pad must be placed on
the NAIL side.
CAUTION: Replace the attachment wrap whenever the adhesive does not adequately adhere the sensor to the patient or
when the attachment wrap becomes soiled from excessive use.
NOTE: Clip the clothing clip to the patient gown to reduce cable pulling on the measuring site.
Clean Shield®
Standard
Multisite Wrap
Petite Wrap
Standard
Foam Wrap
Wrap
7
9123F-eIFU-1019