USING THE MANUAL OVERRIDE

In the event of loss of electrical function, the cot is equipped with a manual override to allow manual operation of the
product until electrical functionality is restored. You can use the red manual back-up release handle to raise or lower
the cot.
The red manual back-up release handle (A) is located along the patient left side of the lower lift bar at the foot end of
the cot as shown in Figure 15.
A
To raise or lower the cot with the manual back-up release handle:
1. Both Operators − Lift the cot during the raise/lower operation to support the weight of the cot at each end.
2. Operator 1 (Foot End) − Pull the manual back-up release handle toward the lift bar. While the manual back-up
release handle is pulled, raise or lower the cot to the desired position and then release the handle to lock the cot
into position.
Note:
• The operators must lift the cot weight slightly off of the wheels to use the manual extend or retract while a patient
is on the cot.
• Activation of the manual back-up release handle may cause the cot to drop slowly if less than 7 stones (45 kg)
(100 pounds) are on the cot.