Cot Operation

Manual Override Operation

Figure 51 - Manual Release Handle
In the event of loss of electrical function, the ambulance cot is equipped with a manual override to allow manual operation
of the product until electrical functionality is restored.
The red manual release lever (A) is located along the patient left side of the lower lift bar at the foot end of the cot.
To raise or lower the cot with the manual release:
Both Operators - lift the cot during the raise/lower operation to support the weight of the cot.
With the ambulance cot supported by an operator at each end, the operator at the foot-end must pull the release handle
towards the lift bar. While the release handle is pulled, raise or lower the ambulance cot to the desired position and then
release the handle to lock ambulance cot into position.
NOTE
The operators must lift the cot weight slightly off the wheels to use the manual extend or retract while a patient is on the
cot.
Activation of the manual release may cause the ambulance cot to drop slowly if less than 40 pounds (18 kg) are on the
cot, which may cause injury to the patient or operator.