Error and Warning Messages
en
Message in the pump display
Communication Error
Calibration Error
Electronics Error
Motor Error
Sensor Error
Defective Key
Overpressure Check Stopcock
Check Flow Path Stopcock, Clamps 3 audible warning signals
Overpressure Open Stopcock
Deficit Limit Reached
Deficit Limit Exceeded
Scale Test
Check Scale Connection
Scale Overloaded Check Scale
Prime Fail-Open Stopcock, Clamps 3 audible warning signals
High Fluid Loss Check Leakage
Low Vac Failed Use Alternative
High Vac Failed Use Alternative
Vac Systems Out Use Alternative
Container Change, Press Resume
Prime Successful Close Stopcock
System OK
Check Tube Set Installation
Tube Set Over Usage Limits
54 / 412
11
Error and Warning Messages
The messages appear in the pump display and audible warning signals are heard.
Audible warning signals indicating a warning or operating message are sounded
several times.
audible warning signals
Method
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
Pressure has exceeded the 200 mmHg safety limit and must
5 audible warning signals,
be reduced. The most common cause is a closed hystero-
constantly repeating until
scope stopcock while the pump operates at the highest flow
pressure is reduced
rate.
Flow path is blocked. Check that clamps and hysteroscope
stopcock are open. Make sure the tube set is not blocked.
Most commonly triggered when hysteroscope stopcock is
closed while pump is operating at peak flow rate. Open hys-
3 audible warning signals
teroscope stopcock or remove other closure to relieve pres-
sure.
3 audible warning signals
Actions are at the discretion of the physician.
Actions are at the discretion of the physician. If necessary,
3 audible warning signals
perform manual deficit control.
The scale test is active. Please proceed as described in -Ref to
3 audible warning signals
Scale Test Description-.
3 audible warning signals,
Check scale connection. Reconnect scale, restart device. If
constantly repeating until
the message reappears, contact Hologic.
connection is reestablished
Weight on the container scale exceeds 25 kg (55 lbs). The
3 audible warning signals
weight must be reduced. System function will resume when
the excess weight has been removed.
Check clamp(s) and hysteroscope inflow stopcock are open.
Press the Prime button to restart.
Actions are at the discretion of the physician. If necessary,
3 audible warning signals
perform manual deficit control.
3 audible warning signals,
An alternative low vacuum source must be used in order to
constantly
repeating
as
continue the procedure. Contact Hologic Technical Support.
long as the error occurs
3 audible warning signals,
An alternative high vacuum source must be used in order to
constantly
repeating
as
continue the procedure. Contact Hologic Technical Support.
long as the error occurs
3 audible warning signals,
An alternative low vacuum source must be used in order to
constantly
repeating
as
continue the procedure. Contact Hologic Technical Support.
long as the error occurs
The system has recognized an ongoing container change
3 audible warning signals
and is waiting for user action to proceed.
The lumen calibration was finished successful. Please close
3 audible warning signals
the Stopcock at the instrument and proceed by inserting the
instrument.
The system check was finished successful. All system com-
1 audible warning signal
ponents are ready to use, no error detected.
Remove and re-insert tube set. If the message reappears, in-
1 audible warning signal
sert a new tube set.
Tube set detection indicates that the tube set has been used
1 audible warning signal
already. Insert new tube set.