Error and Warning Messages

en
Message in the pump display
Check Tube Set Installation
Tube Set Over Usage Limits
Check Flow Path, Stopcock, Clamps 3 audible warning signals
Incorrect Tube Set
Pump Paused, Press Resume
Overpressure Open Stopcock
Overpressure Check Stopcock
Maximum Pressure
Deficit Limit Reached
Deficit Limit Exceeded
Pressure Threshold
Check Scale Connection
Remove Tube Set for System Check 1 audible warning signal
Prime Fail-Open Stopcock, Clamps 3 audible warning signals
High Fluid Loss Check Leakage
Scale Overloaded/Check Scale
Communication Error
Calibration Error
Sensor Error
Motor Error
Low Vac Failed Use Alternative
High Vac Failed Use Alternative
Vac Systems Out Use Alternative
48 / 378
10
Error and Warning Messages
The messages appear in the pump display and audible warning signals are heard.
Audible warning signals indicating a warning or operating message are sounded
several times.
audible warning signals
Method
Remove and re-insert tube set. If the message reappears, in-
1 audible warning signal
sert a new tube set.
Tube set detection indicates that the tube set has been used
1 audible warning signal
already. Insert new tube set.
Flow path is blocked. Check that bag clamps and hystero-
scope stopcock are open. Make sure the tube set is not
blocked.
Replace tube set. The tube set does not match the type ap-
1 audible warning signal
proved for the Aquilex® Fluid Control System .
Pause/Resume button has been activated. Press Pause/Re-
1 audible warning signal
sume button again to resume surgery.
Most commonly triggered when hysteroscope stopcock is
closed while pump is operating at peak flow rate. Open hys-
3 audible warning signals
teroscope stopcock or remove other closure to relieve pres-
sure.
Pressure has exceeded the 200 mmHg safety limit and must
be reduced. The most common cause is a closed hystero-
5 audible warning signals,
scope stopcock while the pump operates at the highest flow
constantly repeating until
rate.
pressure is reduced
Open inflow stopcock on hysteroscope or remove blockage
clamping off the inflow tube set.
The message "Maximum Pressure" is depicted when the in-
No audible signal
trauterine pressure is greater than 150 mmHg.
3 audible warning signals
Actions are at the discretion of the physician.
Actions are at the discretion of the physician. If necessary,
3 audible warning signals
perform manual deficit control.
For a hysteroscopy, pressures above 100  mmHg are usually
1 audible warning signal
not necessary. Careful monitoring of the fluid deficit is re-
commended.
Check scale connection. Reconnect scale, restart device. If
3 audible warning signals
the message reappears, contact Hologic.
Make sure all tube sets are removed from the roller wheel
during the system test. Remove tube set and wait until you
hear an audible warning signal and the message "Insert
Tube Set" appears.
Check bag clamp(s) and hysteroscope inflow stopcock are
open. Press the "Prime" button to restart.
Actions are at the discretion of the physician. If necessary,
3 audible warning signals
perform manual deficit control.
Weight on the container scale exceeds 25 kg (55 lbs) or
weight on the bag scale exceeds 12 kg (26.5 lbs). The weight
3 audible warning signals
must be reduced. System function will resume when the ex-
cess weight has been removed.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
5 audible warning signals
Contact Hologic Technical Support.
An alternative low vacuum source must be used in order to
3 audible warning signals
continue the procedure. Contact Hologic Technical Support.
An alternative high vacuum source must be used in order to
3 audible warning signals
continue the procedure. Contact Hologic Technical Support.
An alternative low vacuum source must be used in order to
3 audible warning signals
continue the procedure. Contact Hologic Technical Support.