• Keep the hysteroscope at the height of the patient and above the drape to en-
sure the fluid can be collected. Do not insert the hysteroscope into the uterus
at this time.
• Press the Prime (17) button (Fig. Front of irrigation pump unit [} 25]).
• Pump is running to carry out the automatic lumen calibration. The pump dis-
play depicts Calibration Running.
• Three audible warning signals are heard once the automatic lumen calibration
is finished.
• The pump display depicts Prime Successful Close Stopcock for 5 seconds, fol-
lowed by Pump Operating.
• Close the stopcock for the hysteroscope inflow to stop the inflow.
5.16 Total Inflow Volume Displayed
If a manual check of the fluid deficit is desired, the total fluid volume supplied from
the fluid bags can be obtained by simultaneously pressing and holding both the up
and down arrows ((10) and (11) Fig. Front of irrigation pump unit [} 25]). The num-
ber in the deficit display is the total inflow fluid volume in ml. Once one or both of
these buttons is released, the deficit display will return to the fluid deficit value.
5.17 Turning System Off
Press the ON/OFF switch to turn pump off. The displays and indicators are no
longer illuminated.
WARNING!
Disconnect the power cord
Pressing the ON/OFF switch does not disconnect the system from the wall power
outlet. This requires pulling the power cord located in the rear of the system.
System Operation
Total inflow volume displayed
Turning off
en
41 / 412