Purpose
en
20 / 412
CAUTION!
Continuous operation
After 24 hours of continuous operation, a device self-test must be carried out.
Switch device off and on again.
CAUTION!
Accessories
To ensure compliance with the requirements of IEC 60601-1-2 in the current ver-
sion, the device Aquilex® Fluid Control System must be used only with the ac-
cessories listed in Chapter Accessory List [} 60].
CAUTION!
Not to be used with a defibrillator
The device may not be used in conjunction with a defibrillator since it is not
equipped with corresponding safety elements. The manufacturer accepts no liabil-
ity in this case for ensuing damage.
CAUTION!
Mains Power Cable
Any power cables employed by the user that are not provided by the manufacturer
must meet the safety requirements of the national standards in the respective cur-
rent valid version.
CAUTION!
Ventilation of the device
• Avoid device overheating.
• Ensure free air circulation especially to the bottom and rear of the device (rear
panel distance of at least 10 cm).
CAUTION!
Modifying the system
Modifying the system and/or its accessories is not permitted.
CAUTION!
Electrical interference
(see Chapter Electromagnetic Compatibility [} 56]): Electrical interference with
other devices or instruments was considered when developing this system and
none was detected during testing. However, if you still detect or suspect such inter-
ference, please follow these suggestions:
• Move the Aquilex® System, the other device, or both devices to a different loc-
ation.
• Increase distance between devices used.
• Consult an electro-medical expert.