The table below shows Arms (Accuracy Root Mean Square) values measured using the M-LNCS/LNCS Sensor with Masimo
SET Oximetry Technology in a clinical study.
SaO
versus error (SpO
2
2
10.0
5.0
0.0
-5.0
-10.0
70
10.0
5.0
0.0
-5.0
-10.0
70
PERFORMANCE SPECIFICATIONS
M-LNCS/LNCS - Adtx/Pdtx
Range
A
rms
90-100%
1.64%
80-90%
1.07%
70-80%
1.55%
– SaO
) with linear regression fit and upper 95% and lower 95% limits of agreement
2
Adt Masimo Saturation Accuracy
M-LNCS/LNCS - Adtx/Pdtx
75
Neo Masimo Saturation Accuracy
M-LNCS/LNCS - Inf/Neo/NeoPt
75
M-LNCS/LNCS - Inf/Neo
90-100%
80
85
SaO2
80
85
SaO2
117
Range
A
rms
1.85%
80-90%
1.44%
70-80%
0.89%
90
90
95
100
95
100