53210A Math, Graphing, and Data Logging
6
Viewing Trend Charts
At power-on or following a reset (
) or instrument preset (Preset), numeric
*RST
data is displayed. The display can be changed to a trend chart by changing the
display mode using the command:
DISPlay[:WINDow]:MODE {NUMeric|HISTogram|TCHart}
DISPlay[:WINDow]:MODE?
(query form)
The display mode (
) is the only trend chart parameter that can be set
TCHart
programmatically (i.e. using SCPI commands). All other trend chart parameters
are set using the front panel.
Pressing the Graph key followed by the Trend Chart softkey starts the trend chart
from the front panel.
Trend charts correspond to all readings in reading memory for a given trigger
count (
) and sample count (
). If triggering is
TRIGger:COUNt
SAMPle:COUNt
continuous (i.e. internal), the trend chart continually updates.
Trend Chart Configuration
Figure 6-9
shows the softkey menus associated with trend chart configuration and
control of the trend window.
Keysight 53210A User's Guide
179