4
53210A Input Signal Conditioning
Threshold Level and Sensitivity
The threshold level is the trigger (input) level at which the counter begins the
signal count (i.e. measurement). This level is the center of the hysteresis band -
the band which represents counter sensitivity. For a count to occur, the signal
must cross the upper and lower limits of the band in opposite directions
(polarities).
signal. The dynamic range of the input signal is indicated on the front panel.
trigger point
V
U
V
C
V
L
positive slope
negative slope
Figure 4-4
112
Figure 4-4
identifies these characteristics and conditions of the input
reset point
INPut[1]:SLOPe
INPut[1]:LEVel:PTPeak?
Input Signal Threshold Level and Sensitivity.
INPut[1]:LEVel
SYSTem:ALEVel:FREQuency
INPut[1]:LEVel:AUTO
INPut[1]:LEVel:RELative
INPut[1]:NREJect
INPut[1]:LEVel:MAXimum?
0
front panel dynamic range indicator
INPut[1]:LEVel:MINimum?
threshold level
Hysteresis band
(p-p sensitivity)
Keysight 53210A User's Guide