Input Range
The signal operating ranges (input ranges) of the 53210A counter are ± 5.0V, ±
50V,and ± 500V and are dependent on the probe factor. The range is set using the
command:
INPut[1]:RANGe {<range>|MINimum|MAXimum|DEFault}
INPut[1]:RANGe? [{MINimum|MAXimum|DEFault}]
When a 1:1 probe factor is selected (see
ranges are 5.0V and 50.0V. When a 10:1 probe factor is selected the available
ranges are 50.0V and 500.0V.
CONFigure
(
*RST
Using Auto Scale
For input signals greater than 100 Hz, Auto Scale detects the signal on the input
channel and sets the range to 5V or 50V depending on the amplitude. Auto Scale
is not the same as an instrument preset which sets a default range of 5V.
Range Example
//period measurement of expected 10 MHz signal,
//maximum resolution, use channel 1
CONF:PER 0.1E-6,MAX,(@1)
Keysight 53210A User's Guide
and
MEASure
) or a front panel preset (Preset), the range is set to 5.0V.
INP:IMP 1E6
//set impedance to 1 Mohm
INP:RANG 50
//set 50V range
53210A Input Signal Conditioning
"Setting the Probe
do not change the input range setting. Following a reset
(query form)
Factor"), the available
107