3
53210A Measurements
//configure counter for frequency measurements
//change parameter values from those set by CONFigure
CONF:FREQ 1.0E6, (@2)
INIT
Using CONFigure?
The following command:
CONFigure?
returns the measurement function configured by the last
command sent to the instrument. Sending
generates a settings conflict error until
For example:
CONF:FREQ 1.0E6, (@2)
CONF?
returns (including quotes):
"FREQ +1.00000000000000E+006,+1.00000000000000E-004,(@2)"
which includes the function, expected value, (calculated) resolution, and channel.
If a channel number is not specified in the
channel is included in the return string.
92
TRIG:SOUR EXT
TRIG:SLOP POS
TRIG:COUN 2
SAMP:COUN 5
SENS:FREQ:GATE:TIME 0.005
SENS:FREQ:GATE:SOUR TIME
CONFigure
after power is cycled
CONFigure?
or
CONFigure
MEASure?
or
CONFigure
MEASure
Keysight 53210A User's Guide
or
MEASure?
is sent.
command, no