Télécharger Imprimer la page

Keysight 53210A Mode D'emploi page 224

Publicité

Les langues disponibles

Les langues disponibles

8
Instrument Status
The Event Register is read with the command:
STATus:QUEStionable[:EVENt]?
Unlike the Condition Register, the bits in the Event Register remain set following
the condition that occurred. The bits are cleared by reading the register.
Enable Register
The Enable register specifies which bits in the Event register can generate a
summary bit which is subsequently used to generate a service request.
Bits within the Event Register are enabled with the command:
STATus:QUEStionable:ENABle <enable>
STATus:QUEStionable:ENABle?
Standard Operation Register Group
The counter's Standard Operation Register Group monitors operational condions
within the counter. These include:
224
: decimal value corresponding to the binary-weighted sum of the bits in
enable
the register. For example, to enable the bit monitoring possible time base
errors:
STAT:QUES:ENAB 1024
– calibration in progress
– battery operation
– battery charge status
– measurement in progress
– waiting for trigger
– internal reference oscillator in use
– remote interface lock and remote interface errors
– configuration change
– reading memory threshold reached
.
Keysight 53210A User's Guide