3
53210A Measurements
The commands used to make frequency measurements are:
MEASure:FREQuency? [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
CONFigure:FREQuency [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
Frequency Measurement Examples
// using MEASure? - measure 20 MHz signal with 0.1 Hz
// resolution
*RST
MEAS:FREQ? 20E6, 0.1, (@1)
-----------------------------------------------------------------------
// using CONFigure - measure 20 MHz signal with
// 0.1 Hz resolution, take 10 readings
*RST
CONF:FREQ 20E6, 0.1, (@1)
READ?
Notes
1 See
94
is the expected input signal frequency.
expected
measurement resolution in hertz. The parameter ranges are:
(channel 1): 0.1 Hz - 350 MHz (default = 10 MHz)
expected
(channel 2 Option 106): 100 MHz - 6.0 GHz (default = 500 MHz)
expected
(channel 2 Option 115): 300 MHz - 15 GHz (default = 500 MHz)
expected
(all channels): 1.0E-15*
resolution
(default
resolution
is counter channel 1 or 2 specified as
channel
SAMP:COUN 10
Chapter 5
for information on triggering and number of readings per
trigger (sample count).
expected
corresponds to a 0.1s gate time)
// reset to start from known state
// reset to start from known state
// take 10 readings
is the desired
resolution
to 1.0E-5*
expected
or
.
(@1)
(@2)