53210A Math, Graphing, and Data Logging
CALCulate[1]:AVERage:STATe {OFF|ON}
CALCulate[1]:AVERage[:STATe]?
ON
includes: mean, standard deviation, maximum value, minimum value, and average
peak-to-peak value.
OFF
Following a reset (
computations are disabled.
Reading Count
Statistical calculations start when Math and the Statistics function are enabled.
The number of readings on which a given set of statistics are based is shown as
the Count value
CALCulate[1]:AVERage:COUNt:CURRent?
The count can be read at any time after the measurements have been initiated (i.e.
INITiate:IMMediate
Following a reset (
'1'.
Mean, Standard Deviation, Minimum Value, Maximum Value
The mean (average), standard deviation, minimum value, and maximum value for
the current reading count can be determined with the command:
CALCulate[1]:AVERage:ALL?
If the readings have been scaled (see
on the scaled readings.
158
- enables statistical calculations on the readings as they occur. Statistical data
- disables statistics computation.
*RST
(Figure
,
READ?
*RST
) or instrument preset (
6-3). The count can also be queried with the command:
,
).
MEASure?
) or instrument preset (
"Scaling
), statistical
SYSTem:PRESet
), the count is set to
SYSTem:PRESet
Functions"), the statistics are based
Keysight 53210A User's Guide
(query form)