Table A-1
53210A Error Message Descriptions.
Code
Message
+820
Model and Serial Numbers not restored
Controller and measurement board
+821
model numbers do not match
Controller and measurement board serial
+822
numbers do not match
Calibration Errors
Calibration error;
+701
security defeated
Calibration error;
+702
calibration memory is secured
Calibration error;
+703
secure code provided was invalid
Calibration error;
+704
secure code too long
Calibration error;
+705
calibration aborted
Calibration error;
+706
provided value out of range
Calibration error;
+707
computed correction factor out of range
Calibration error;
+711
calibration string too long
+712
Calibration failed
Channel 2 calibration signal not
+713
detected
Keysight 53210A User's Guide
Description
Following replacement of the P500 processor (controller) board or
main measurement board, the model number and/or serial
number do not match or have not been re-stored. Re-store the
model number and serial number as prompted from the front
panel. Refer to the 53210A Service Guide for additional
information.
Shorting the calibration secure jumper (CAL ENABLE) when
turning ON the instrument will cause this error to occur indicating
the security password has been overwritten. See "Resetting the
Security Code to a Null" in the 53210A Service Guide for details.
A calibration cannot be performed when calibration memory is
secured. See "To Unlock The Counter For Calibration" and "To
Lock the Counter" in Chapter 2 of the 53210A Service Guide for
more information. Use the CAL:SEC:STAT ON command to enter
the security code using the remote interface.
The security code specified was invalid.
The security code specified is greater than 12 characters.
An in-progress calibration was aborted (stopped) by the user.
A calibration value was entered that was outside the expected
range of valid inputs.
If this error occurs, cycle power and run the instrument self-test.
Verify the calibration setup (instrument settings, connections) and
recalibrate the counter. Contact Keysight if there is a self-test
failure or the unit fails to calibrate.
The message to be stored in calibration memory is greater than 40
characters.
A hardware failure occurred. Contact Keysight.
The calibration source signal is not connected to channel 2.
53210A Error Messages
A
237