Using the Scale Functions
The counter scaling functions include null, percent change (PCT), part per million
change (PPM), part per billion change (PPB), and scale (Mx-B). The function is
selected with the command:
CALCulate[1]:SCALe:FUNCtion {NULL|PCT|PPM|PPB|SCALe}
CALCulate[1]:SCALe:FUNCtion?
NULL
reference value. (See
reference).
PCT
the measurement and the reference value, expressed in percentage.
PPM
between the measurement and the reference value, expressed in parts per million.
PPB
between the measurement and the reference value, expressed in parts per billion.
SCALe
by the gain value M (
(
CALCulate1:SCALe:OFFSet
measurement will first be inverted (1/x), resulting in a M/x-B operation.
Keysight 53210A User's Guide
- performs a null operation. The result is the measurement minus the
"Scale Reference
- performs a percent change operation. The result is the difference between
- performs a parts per million change operation. The result is the difference
- performs a parts per billion change operation. The result is the difference
- performs an Mx-B operation. The result is the measurement (x) multiplied
CALCulate1:SCALe:GAIN
53210A Math, Graphing, and Data Logging
Value" for information on setting the
command) minus the offset value B
). If
CALCulate1:SCALe:INVert
(query form)
is enabled (
), the
On
6
151