Triggering and Gating
5
Idle State
Counter configuration generally occurs while the instrument is in the idle state
(Figure
5-2). This includes configuration of the system trigger.
As shown, the counter is placed in the idle state by either of the following the
commands:
- aborts a measurement in progress.
ABORt
- resets the counter to its factory default values.
*RST
The counter also returns to the idle state after the total number of measurements
determined by the
and
settings are complete (i.e. #
TRIGger:COUNt
SAMPle:COUNt
of measurements =
x
).
TRIGger:COUNt
SAMPle:COUNt
System Trigger Source
\
The system trigger is configured using the commands within the
and
TRIGger
subsystems.
Figure 5-2
show the sequence in which the commands are
SAMPle
commonly used.
The system trigger source which starts the trigger/gate cycle is set with the
command:
TRIGger:SOURce {IMMediate|EXTernal|BUS}
TRIGger:SOURce?
(query form)
Keysight 53210A User's Guide
129