A
53210A Error Messages
Table A-1
53210A Error Message Descriptions.
Code
Message
Channel 2 calibration signal power level
+714
error
Channel 2 calibration signal frequency
+715
error
Channel 2 calibration signal
+716
is not CW
Channel 2 calibration
+717
timeout occurred
Auto-calibration failed;
+720
input signal detected
Calibration data lost:
+742
corrections
Calibration memory
+748
write failure
+750
Calibration data not restored
Sel f-Test Errors
Sel f-Test failed:
+901
auto-calibration failure
Sel f-Test failed:
+903
real-time clock setting lost
238
Description
The power level of the calibration signal was outside the expected
range of the specified level.
The frequency of the calibration signal was outside the expect
range of the specified value.
The Channel 2 calibration signal must be a continuous wave (CW)
and not a pulsed signal.
Calibration at the specified power level did not complete within
the ~ 1.2 seconds allowed. Perform a self-test to verify operation
of the Channel 2 board and re-check the parameters of the input
signal.
No signal should be present on the input channel during an
auto-cal.
The instrument's calibration data was lost due to a firmware
update that may have made the current calibration data invalid, or
due to a hardware failure. Recalibrate the instrument.
An error occurred as calibration data was being written to the
counter's internal flash memory. Run the self-test to verify counter
operation and repeat the calibration procedure.
Following repair, calibration data was not restored as prompted
from the front panel.
A failure occurred during the auto-calibration portion of the
self-test. Repeat the self-test. If the auto-calibration failure
remains there is an issue with the instrument hardware. Contact
Keysight.
Reset the instrument clock using the 'Utility' -> 'System Setup' ->
'Date/Time' keys, or using the SYSTem:TIME or SYSTEM:DATE
commands. Re-run the self-test.