Télécharger Imprimer la page

Keysight 53210A Mode D'emploi page 195

Publicité

Les langues disponibles

Les langues disponibles

Table 7-2
INITiate:IMMediate
MMEMORY:STORe
Front Panel Display
Counter measurements taken from the front panel or from an IO interface appear
in the display as indicated in
features.
Reading Memory
In addition to being displayed, all counter readings are stored in the counter's
volatile read ing memory
Reading memory has the following attributes:
1 reading memory is cleared after a function change, during execution of the
Keysight 53210A User's Guide
Commands Affecting Data Flow
Locations
Command
Accessed
Front Panel
Read?
Reading Memory
Output Buffer
Front Panel
Reading Memory
Reading Memory
FETCh?
Output Buffer
Reading Memory
R?
Output Buffer
Reading Memory
DATA:REMove?
Output Buffer
Reading Memory
DATA:LAST?
Output Buffer
Reading Memory
Internal Flash
:DATA
External USB
,
READ?
INITiate:IMMediate
following an instrument reset (
Read ings
Transferred
All
None
All
All or
specified count
Specified Count
1
All
Figure
7-1. See
Chapter 1
(Figure
7-1). Memory capacity is 1,000,000 readings.
, or
command, after cycling power, or
MEASure?
) or preset (
*RST
Formats and Data Flow
Read ing Memory
Allowed during
Cleared
Measurement
No
Yes
No
Yes
Yes
No
No
for an overview of display
).
SYSTem:PRESet
7
No
No
No
Yes
Yes
Yes
No
195

Publicité

loading