6
53210A Math, Graphing, and Data Logging
A reset (
function disables the
Smoothing Data
Measurements can be "smoothed" and scaled prior to performing math
operations on the incoming data.
To red uce random noise, a moving average (boxcar) filter can be inserted into the
data path
(subset) to average, the following commands are used:
CALCulate[1]:SMOothing[:STATe] {OFF|ON}
CALCulate[1]:SMOothing[:STATe]?
CALCulate[1]:SMOothing:RESPonse {SLOW|MEDium|FAST}
CALCulate[1]:SMOothing:RESPonse?
ON
filter. A reset (
148
), instrument preset (
*RST
CALCulate1
(Figure
6-1). To enable the filter and specify the number of readings
enables the moving average filter - placing it in the data path.
) or instrument preset (
*RST
), or changing the measurement
SYSTem:PRESet
subsystem.
SYSTem:PRESet
Keysight 53210A User's Guide
(query form)
(query form)
disables the
OFF
) disables the filter.