Table A-1
53210A Error Message Descriptions.
Code
Message
-350
Error queue overflow
Query Errors
-410
Query IINTERRUPTED
-420
Query UNTERMINATED
-430
Query DEADLOCKED
Query UNTERMINATED after indefinite
-440
response
Instrument Errors
+100
Network Error
+110
LXI mDNS Error
Not able to execute while instrument is
+263
measuring
+291
Not able to recall state: it is empty
+292
State file size error
+293
State file corrupt
+294
Preference file size error
+295
Preference file corrupt
Input termination protection relay
+301
opened
Keysight 53210A User's Guide
Description
An error has occurred; however, the instrument error queue is full
and the error was not recorded.
The counter received a command before it was finished
responding to a previous query.
The computer attempts to read a query response from the counter
after having issued an incomplete query command.
The counter's input and output buffers are full and the instrument
cannot continue.
A query command was received in the same string after a query
(e.g. *IDN?) requesting an indefinite length response.
A LAN network error has occurred or there is a possible hardware
failure. Run the instrument self-test to determine if a hardware
failure exists. If so, contact Keysight.
A command was received while a measurement is in progress
The state storage location (0-4) specified by the *RCL command is
empty
Attempted to load a .sta (state) file that is too large.
The .sta (state) file specified does not contain valid instrument
state information.
Attempted to load a .prf (preference) file that is too large.
The .prf (preference) file specified does not contain valid
user-preference setting information.
An input signal greater than or equal to + 10Vp is present on a
channel. The input impedance is set to 1 MΩ. Removing or
reducing the signal below the damage level and pressing the
flashing (channel) key or sending INPut:PROTection:CLEar resets
the protection relay.
53210A Error Messages
A
235