Télécharger Imprimer la page

Keysight 53210A Mode D'emploi page 146

Publicité

Les langues disponibles

Les langues disponibles

6
53210A Math, Graphing, and Data Logging
Math Functions
The 553210A counter math functions include smoothing, null/scale, statistics,
and limit checking.
The math functions are enabled at two levels: 1) the
whole is enabled, and 2) the individual math operation is enabled. Math
operations begin once the measurements are started either by
INITiate:IMMediate
front panel (with internal triggering). Readings are processed real time and
displayed, stored in reading memory, and sent to the output buffer (they are not
post-processed from reading memory).
The number of readings processed for a given trigger cycle
"Triggering and
the number of readings per trigger (
sample count set by each command is '1'. Each time the counter is initiated
reading memory is cleared and a new set of math-processed data is generated.
From the front panel, readings are processed continually if the trigger source
remains at its default
The math and graphics functions covered in this chapter are enabled individually
(Figure
performed on the same set of readings.
The SCPI commands and parameters listed throughout this chapter are provided
NOTE
as a guide to how the counter is operated programmatically. The commands are
described in detail in the Programmer's Reference.
146
Figure 6-1
or
READ?
Gating") is determined by the trigger count (
INTernal
6-1). However, multiple operations may be enabled at the same time and
is a description of how the functions are enabled.
, or as soon as the operation is selected from the
). The default trigger count and
SAMPle:COUNt
setting.
subsystem as a
CALCulate[1]
(Chapter 5,
TRIGger:COUNt
Keysight 53210A User's Guide
) and