CALCulate1:SMOothing:STATe ON
CALCulate1:SMOothing
(moving average)
measurement
data
CALCulate1:STATe ON
readings initiated by INITiate:IMMediate are sent to reading memory.
readings initiated by READ? are sent to reading memory and to the output buffer
Figure 6-1
Enabling the CALCulate1 Subsystem
Individual math operations require the
math operation to be enabled before the operation is performed. The command
used to enable the
CALCulate[1][:STATe] {OFF|ON}
CALCulate[1][:STATe]?
ON
d isabled, measurement data is sent directly to the output buffer and/or reading
memory with no math operations performed, regardless of any math operations
enabled.
Keysight 53210A User's Guide
CALCulate1:SCALe: STATe ON
CALCulate1:SCALe
(scaling and offset)
Enabling the Math Operations.
CALCulate1
enables the subsystem,
53210A Math, Graphing, and Data Logging
CALCulate1:AVERage:STATe ON
CALCulate1:AVERage
(statistics)
CALCulate1:LIMit:STATe ON
CALCulate1:LIMit
(limit checking)
CALCulate2:TRANsform:HISTogram:STATe ON
CALCulate2:TRANsform:HISTogram
(histograms)
CALCculate1
subsystem is:
disables the subsystem. With the subsystem
OFF
INITiate:IMMediate
reading memory
READ?
subsystem and the specific
(query form)
6
output
buffer
147