53210A Error Messages
Table A-1
53210A Error Message Descriptions.
Code
Message
Device Specific Errors
System Error;
-310
internal software error
System Error;
-310
failed to erase calibration data in PIC
EEProm
System Error;
-310
failed to erase system information in PIC
EEProm
System Error;
-310
I2C Comms Failure
Calibration memory lost;
-313
memory corruption detected
Calibration memory lost;
-313
due to firmware revision change
Save/recall memory lost;
-314
memory corruption detected
Save/recall memory lost;
-314
due to firmware revision change
Configuration memory lost;
-315
memory corruption detected
Configuration memory lost;
-315
due to firmware revision change
-330
Sel f-test failed
234
Description
At power-on, the firmware is unable to obtain information from the
instrument's non-volatile memory to complete initialization. Cycle
power again. Contact Keysight if the error remains.
For system errors related to erasing, reading, or writing calibration
data, verify that the instrument has the latest firmware revision
installed. Firmware updates can be located beginning at the
"Technical Support" tab at www.keysight.com/find/53210A. If the
error condition remains, contact Keysight.
For system errors related to erasing, reading, or writing system
information, verify that the instrument has the latest firmware
revision installed. Firmware updates can be located beginning at
the "Technical Support" tab at www.keysight.com/find/53210A. If
the error condition remains, contact Keysight.
I2C related errors if they occur, will occur most often during boot
up when power is cycled. If this type of error occurs cycle power
again. If the error remains, contact Keysight with a complete
description of the error message.
The instrument was unable to reference calibration data in
non-volatile memory. Perform a recalibration.
The instrument's calibration data has been erased following a
instrument firmware update. Recalibration of the instrument is
required.
The instrument state saved by the *SAV? command in non-volatile
(flash) memory has been lost.
The instrument state saved by the *SAV? command in non-volatile
(flash) memory has been lost due to a firmware update.
User-preference settings such as I/O settings, the reference
oscillator setting, measurement time out setting,... have been lost.
User-preference settings such as I/O settings, the reference
oscillator setting, measurement time out setting,... have been lost
due to a firmware update.
See the appended message for more information.
Keysight 53210A User's Guide