Table 5-2
Setting the Gate Time
A measurement gate time can be specified directly with the command:
[SENSe:]FREQuency:GATE:TIME {<time>|MINimum|MAXimum|DEFault}
[SENSe:]FREQuency:GATE:TIME? {MINimum|MAXimum|DEFault}
CONFigure
time according to their optional expected value and resolution parameters.
Specifying these parameters or using default values with these commands
simplifies counter programming. However, setting/changing the gate time
separate from the
gate time while preserving other configuration settings.
Keysight 53210A User's Guide
Resolution and Gate Time.
Gate Time (53210A)
(T
= 100 ps)
ss
1 msec
10 msec
100 msec
1 sec
10 sec
100 sec
1000 sec
and
MEASure
CONFigure
Resolution (Digits)
Reciprocal
7
8
9
10
11
12
13
automatically set the gate source to
/
command allows you to change only the
MEASure
Triggering and Gating
Calculating Digits and Gate Time
Digits as a function of gate time:
Digits = Log
(gate time/T
10
Digits as a function of expected value
(CONFigure, MEASure)
Digits = Log
(expected value) -
10
Log
(resolution)
10
Gate time as a function of digits:
gate time = (10^d igits) * T
TIME,
5
)
ss
ss
(query form)
and the gate