Using The Adhesive Patch - Boston Scientific Vercise DBS Mode D'emploi

Masquer les pouces Voir aussi pour Vercise DBS:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Note: The Charging Collar can be placed under or over clothing. You should not
wear tight fitting or heavy clothing over the Charger while charging to allow
air flow around the Charger.

Using the Adhesive Patch

1. When the indicator light is green, remove the Charger from the Base Station.
(The indicator light will go off, regardless of the ready status of the Charger.)
2. Apply the Adhesive Patch to the backside of
the Charger by peeling the clear liner from
the patch and applying the white side with
the blue stripe to the rear of the Charger,
as shown to the right.
3. Remove the skin side beige liner from the
adhesive (only good for one fixation.).
4. For best charging results, make sure the Charger is centered over the Stimulator.
If the Charger is not centered, charging time may increase. See Properly Aligning
the Charger on page 44 to ensure you center the Charger over the Stimulator.
WARNING: Do not put the Charger directly on the skin (e.g., without an Adhesive
Patch). This may result in uncomfortable heating of the skin or a burn.
The Charging System
Physician Manual
91098825-07 Rev A 43 of 333