Note: The Remote Control will terminate the CP Ready state and transition to the
sleep mode if there is no IR signal after 15 minutes.
All buttons are active during CP Ready and pressing any button returns the
Remote Control to the Level screen.
Stimulation may be turned on or off during CP Ready.
Once IR Communication is established, the Remote Control will terminate
communication and transition to the sleep mode if there is no IR activity after
15 minutes.
All buttons are active during IR Communication and pressing any button
returns the Remote Control to the Level screen.
Whenever the Remote Control is re-activated from sleep mode the display will
default to the Level screen.
Selection #3 – Clear Link
When the "Clear Link" option is selected from the Enter Clinician Options menu, you
will immediately be required to enter the clinician's password in order to continue.
The password for the Clear Link option is A-B-C.
To enter the password:
The first character is highlighted when the Enter Password screen opens. To select/
confirm any character and/or move to the next character position, press
through possible characters, use ▲or ▼.
Note: Do not share this password with patients, as they may unknowingly clear the
link between their Remote Control and Stimulator.
When clearing a link:
If the password is entered correctly, the link between the Remote Control and the
"previous" Stimulator is broken immediately, and the Remote Control displays the
Not Linked screen. The Remote Control's programs remain intact. If the password is
entered incorrectly, the process is aborted and the Remote Control will return to the
Enter Clinician Options screen.
Physician Manual
91098825-07 Rev A 56 of 333