Connecting peripheral equipment to the built-in 110 volt auxiliary power outlet
(option)
WARNING
Only use hospital-grade electric equipment consuming 10A or less with the auxiliary power outlet (option). The use
of standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment.
Do not use the 110V auxiliary power outlet (option) for life sustaining equipment.
The 110 volt auxiliary power outlet is a built-in outlet for peripheral equipment. You can find the outlet under the foot end
on the patient right side of the product
www.stryker.com
Operation
(Product illustration on page
1-13).
Figure 1-20: Optional auxiliary outlet
3005-009-005 REV A
English
EN
1-29