Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
EN
WARNING
Always plug the product directly into a grounded, hospital grade wall outlet. You can only achieve grounding
reliability when you use a hospital grade wall outlet. This product is equipped with a hospital-grade plug for
protection against electric shock hazard.
Always use a Stryker supplied interface cable. Use of any other cable may cause the bed to function improperly,
which may result in patient or user injury.
Always associate or map the iBed Locator to the room or location to provide accurate location information. Failure
to properly associate or map the iBed Locator to the room or location will yield incorrect remote information. If you
move an iBed Locator after it has been installed and mapped, you must remap to the new room or location. If the
room or location information is changed after initial installation, you must remap the iBed Locator.
Always apply the brakes when a patient is getting into or out of the product to avoid instability.
Always apply the brakes when the patient is unattended.
Do not apply the brakes to slow or stop the product while it is in motion.
Always lock the siderails in the full up position with the sleep surface horizontal in the lowest position when you
transport the product with a patient.
Always unplug the power cord before you transport the product.
Always release the brakes before you transport the product. Do not transport the product with the brakes applied.
Do not transport the product laterally after you apply the steer lock pedal. The product cannot swivel when you
transport with steer lock.
Use only retractable traction or fracture frames. Failure to use a retractable frame may result in injury to the patient
and/or damage to the equipment.
Always set the siderail position to make sure that a patient is safely in the product.
Always lock the operator control panel and patient control panel when the patient is unattended.
Always keep the siderails outside of the oxygen tent.
Only use hospital-grade electric equipment consuming 10A or less with the auxiliary power outlet (option). The use
of standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment.
Do not use the 110V auxiliary power outlet (option) for life sustaining equipment.
Always lock the control panel when you leave the patient unattended.
Always lock the patient control panel when the patient's condition requires extra safety measures.
Do not use bed exit to replace patient monitoring protocol.
Bed exit is intended only to aid in the detection of a patient exiting the product.
Do not use bed exit with patients who weigh less than 50 lb (23 kg).
Do not use the scale system reading as a reference for medical treatment.
The scale system assists only in the monitoring of the patient's weight variation.
Do not set the scale to zero or weigh the patient when a support surface therapy is active. Motion from the support
surface functions may adversely affect the scale system performance.
Do not use iBed Awareness to replace your patient monitoring protocol.
Do not use iBed Awareness as a lock indicator for siderails. iBed Awareness only detects the position of the
siderails.
1-4
Summary of safety precautions
3005-009-005 REV A
www.stryker.com