CAUTION
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
Always make sure that all persons and equipment are away from the area below and around the Fowler before you
activate the CPR release. The CPR release is for emergency use only.
Always determine the proper use of the restraint straps and restraint strap locations. Improperly adjusted restraint
straps can cause serious injury to a patient. Stryker is not responsible for the type or use of restraint straps on any
Stryker products.
Do not load the IV pole above the safe working load of 40 lb (18 kg).
Always raise the siderails when the litter is in its full down position. This prevents the scale system from weighing a
patient inaccurately.
Always make sure that the siderails are locked before you arm iBed Awareness.
Make sure that you set the desired product parameters before arming iBed Awareness.
Do not use accessories that cover the footboard LED light bars.
Do not clean, disinfect, service, or perform maintenance while the product is in use.
Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and
motors. Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause
unpredictable operation and decreased functionality of any electrical product. Do not return the product to service
until it is completely dry and you have tested for safe operation.
Always clean Velcro
®
after each use. Saturate Velcro with disinfectant and allow disinfectant to evaporate.
Appropriate disinfectant for nylon Velcro should be determined by the hospital.
Always wipe down with clean water (or 70% isopropyl alcohol, if using Virex® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse
and dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could
cause premature degradation of critical components. Failure to follow these disinfecting instructions may void your
warranty.
Do not use quaternary disinfectants formulated with glycol ethers.