The 3005 S3 MedSurg bed is intended for use in the electromagnetic environment specified below. The customer or
the user of the 3005 S3 MedSurg bed should assure that it is used in such an environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Recommended separations distances between portable and mobile RF communication equipment and the 3005
The 3005 S3 MedSurg bed is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the 3005 S3 MedSurg bed can help prevent electromagnetic
interferences by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the 3005 S3 MedSurg bed as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power of
transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
1-64
EMC information
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class A
Class A
Complies
150 kHz to 80 MHz
W
D=(1.2) (√P)
0.01
0.1
1
10
100
Electromagnetic environment
The 3005 S3 MedSurg bed uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
The 3005 S3 MedSurg bed is suitable for use in all
establishments other than domestic and those
directly connected to the public low voltage power
supply network that supplies buildings used for
domestic purposes.
S3 MedSurg bed
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
D=(1.2) (√P )
1.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
3005-009-005 REV A
m
800 MHz to 2.5 GHz
D=(2.3) (√P )
0.23
0.73
2.3
7.3
23