System requirements and recommendations for iBed Wireless (option)
EN
(Continued)
Customer WLAN environment
Channel utilization
Signal strength range (minimum)
Minimum SNR
Priority queuing
Client exclusion
Client load balancing
Max number of SSIDs
Authentication timeouts
Note: A transmit power asymmetry problem may arise at the edges of virtual cell coverage if an APs transmit power is
higher than the Stryker Wireless Client device (~6 mW 2.4 GHz or 12 mW 5 GHz). The received signal strength indicator
(RSSI) of the Stryker iBed Wireless Client on the AP must be verified. The device should never drop below an RSSI of
-75 dBm on the AP.
1-12
Introduction
Consistently less than 30%
2.4 GHz: -67dBm +0/-8dBm
5 GHz: -67dBm +0/-8dBm
Minimum 20dB
Prioritized over best effort traffic
Disabled
Disabled
5
Add session timeout of at least 24
hours
3005-009-005 REV A
Required
Recommended
Yes
Yes
Recommended
Recommended
Recommended
Recommended
Recommended
www.stryker.com