Footboard control panel - Chaperone bed exit (option)
EN
WARNING
•
Do not use bed exit to replace patient monitoring protocol.
•
Bed exit is intended only to aid in the detection of a patient exiting the product.
•
Do not use bed exit with patients who weigh less than 50 lb (23 kg).
A
Zone
Arm/Disarm
B
Note: The Bed Exit footboard LED indicator illuminates when you arm Chaperone Bed Exit
Arming or disarming Chaperone bed exit (option)
When armed, Chaperone bed exit monitors the patient's position on the product.
Note: A notification appears if there is not enough weight on the product to arm bed exit.
To arm bed exit:
1.
Set the scale to zero. See
Note: If you do not set the scale to zero before placing a patient on the product, bed exit may not operate properly.
2.
Position the patient on the product.
3.
Press and hold the Arm/Disarm button (A). See
36.
After arming Chaperone Bed Exit, the LED light bars on the footboard illuminate green, the Bed Exit footboard LED
indicator on the footboard LED indicator illuminates, and the selected zone on the footboard control panel illuminates.
If the parameter conditions selected for Chaperone bed exit are changed:
•
LED light bars on the footboard flash amber
•
bed exit indicator LED on the footboard LED indicator flashes
•
sound alarm is triggered
•
bed exit status alert is displayed on the display screen
1-36
Operation
A
B
The patient can move freely, but the alarm sounds when the patient
attempts to leave the product
Arms or disarms bed exit
Setting the scale to zero on page
Footboard control panel - Chaperone bed exit (option) on page 1-
3005-009-005 REV A
(Figure 1-24 on page
1-41.
1-39).