Arming or disarming Chaperone bed exit (option) (Continued)
To disarm bed exit, press and hold the Arm/Disarm button again (A). See
(option) on page
1-36.
Footboard control panel - Chaperone bed exit with zone control (option)
WARNING
Do not use bed exit to replace patient monitoring protocol.
Bed exit is intended only to aid in the detection of a patient exiting the product.
Do not use bed exit with patients who weigh less than 50 lb (23 kg).
D
A
Zone 1
B
Zone 2
C
Zone 3
Arm/Disarm
D
E
Zone
Note: The Bed Exit footboard LED indicator illuminates when you arm Chaperone Bed Exit
Arming or disarming Chaperone bed exit with zone control (option)
When armed, Chaperone bed exit monitors the patient's position on the product.
Note: A notification appears if there is not enough weight on the product to arm bed exit.
www.stryker.com
Operation
Footboard control panel - Chaperone bed exit
B
A
C
Allows the patient to move freely on the bed. Alarms when the patient
moves 50 percent of their body weight out of the designated zone.
Allows for limited movement. Alarms when the patient approaches the
siderail or the foot end of the bed.
Allows minimal movement. Alarms when the patient moves out of the
tightly restricted zone.
Arms or disarms bed exit
Changes the zone
3005-009-005 REV A
E
(Figure 1-24 on page
1-39).