Performance Specifications - Masimo Radius PPG Mode D'emploi

Masquer les pouces Voir aussi pour Radius PPG:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8

PERFORMANCE SPECIFICATIONS

The table below shows A
with Masimo Technology in a clinical study.
See Bland and Altman. Agreement between methods of measurement with multiple observations per individual Journal of Biopharmaceutical Statistics (2007) vol. 17
*
pp. 571-582.
(Accuracy Root Mean Square) values measured using the Radius PPG sensors under no motion,
rms
90–100%
80–90%
70–80%
70–100%
SpO
Upper and Lower Limits of Agreement (LoA)*
2
Upper 95% LoA
Lower 95% LoA
(SpO
-SaO
) vs. (SpO
+SaO
)/2 Bland Altman fit and upper 95% and lower 95% limits of agreement.
2
2
2
2
Radius PPG Sensor
SpO
A
2
rms
1.14 (1,14) %
1.29 (1,29) %
1.41 (1,41) %
1.33 (1,33) %
Actual Value
1.98%
-2.91%
70–100%
Radius PPG No Motion
95