POWER PRO XT ambulance cot is designed to be power washable. The unit may show some signs
of oxidation or discoloration from continuous washing, however, no degradation of the cot's performance characteristics
or functionality will occur due to power washing as long as the proper procedures are followed.
Thoroughly clean the cot once a month. Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow
disinfectant to evaporate. (Appropriate disinfectant for nylon Velcro should be determined by the service.)
Washing Procedure:
Remove the battery! The battery and charger are not immersible or power washable.
Follow the cleaning solution manufacturer's dilution recommendations exactly.
The preferred method Stryker Medical recommends for power washing the POWER PRO XT ambulance cot is with
the standard hospital surgical cart washer or hand held wand unit.

Washing Limitations:

WARNING
Use any appropriate personal safety equipment (goggles, respirator, etc.) to avoid the risk of inhaling contagion. Use of
powerwashing equipment can aerate contamination collected during the use of the cot.
CAUTION
DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE UNIT.
Maximum water temperature should not exceed 180°F/82°C.
Maximum air dry temperature (cart washers) is 240°F/115°C.
Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand held wand is being used to wash the unit, the
pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.
Towel dry all casters and interface points.
Failure to comply with these instructions may invalidate any/all warranties.
Remove the battery before washing the cot.
5

Cleaning

6500-009-005 REV A