General Inspection

The condition of the ambulance cot is the responsibility of the owner. It is important that the ambulance cot is working
properly before the product is put into service. Have a qualified service person use the following list and the operation
instructions to check the ambulance cot before the product is put into service.
Unpack the cartons and check all items for proper operation.
Item
Battery
Unpack batteries and charger
Charge battery according to D
CAUTION
When charging batteries in an ambulance vehicle, locate the charger either in the forward cab or an enclosed compartment
(i.e. cabinet).
Prior to checking the features and condition of the cot, the battery must be charged until the red LED lights continuously
to ensure a full charge. Refer to the D
Once the battery is fully charged, inspect the ambulance cot for the following points:
Item
Battery
Charge spare battery (if necessary) according to D
Install battery into foot-end enclosure, battery indicator operates
Ensure the battery remains firmly secured
Release and remove battery from foot-end enclosure
Reinstall battery into foot-end enclosure
Hydraulics
Inspect motor mount, all fasteners secure
Check cylinder attachments at both ends, all fasteners secure
Inspect main cable, all connections secure
Inspect hoses and cylinder seal for leaks
Electronic Controls Check battery indicator, charged
Extend cot to raised position
Verify "jog" function operating smoothly
Lower to retracted position, cot secures in a mid-height position
(motor does not operate)
Determine and set ambulance vehicle load height
Check high speed retract
Extend cot to full height, no drift
Manual Release
Verify the manual release lever functions properly, adjust accordingly
With the cot empty, check the raise/lower function
With the cot loaded with a minimum of 45 kg,
check the raise/lower function
With the cot loaded with a minimum of 45 kg,
check the load/unload function
www.stryker.com

Product Inspection

Routine
WALT
®
instructions
e
WALT
®
instructions for further charging information.
e
Routine
6500-009-005 REV A
P/N 6500-700-006
WALT
instructions
P/N 6500-700-006
®
e
33 & 54
54
54
54
63
63
63
63
33
51
51
51
18
43
51
47
47
47
44
Page(s)
Page(s)
15

Publicité

Table des Matières
loading

Table des Matières