Unloading the Cot from a Vehicle - Powered Method
Unloading an unoccupied ambulance cot from a vehicle can
be accomplished by a single operator.
WARNING
The one person loading and unloading procedures are for
use only with an empty ambulance cot. Do not use the
procedures when unloading a patient. Injury to the patient or
operator could result.
Disengage the cot from the cot fastener (For more detailed
instructions, reference the ambulance cot fastener
installation/operation manual - Stryker part number
6370-090-010).
Lift the vehicle bumper to the raised position (if possible).
Grasp the cot frame at the foot end. Pull the cot out of the
patient compartment until the safety bar engages the
safety hook. Operator two should verify that the bar
engages the safety hook.
WARNING
Do not pull or lift on the safety bar when unloading the cot.
Damage to the safety bar could result and injury to the
patient or operator could occur.
Depress the extend (+) button to lower the undercarriage to
its fully extended position (Figure 48).
Push the safety bar release lever forward to disengage
the safety bar from the safety hook in the patient
compartment.
Remove the load wheels from the patient compartment of
the vehicle.
CAUTION
When unloading the ambulance cot from the patient
compartment, ensure the caster wheels are safely set on the
ground or damage to the product may occur.
Do not "jog" the cot past the load height while the safety bar
is engaged.