Stryker Power Pro Cot 6500 RUGGED Manuel D'utilisation page 73

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83
Adjusting the Manual Release Cable
Tools:
8mm wrench
10mm wrench
1.
Support the litter so no weight is on the base.
2.
Ensure the cable is intact.
3.
Using a 10mm wrench, loosen the cable lock nut.
4.
Using a 8mm hex wrench, adjust the tension on the manual release cable.
5.
Tighten cable lock nut.
Filling the Reservoir - Use only Mobil Mercon Synthetic Blend Oil (6500-001-293)
Note: Any time you work with the hydraulics you may lose some oil.
Tools:
3/16" Allen wrench
1.
Raise the cot to the full up position.
2.
Ensure the fill port is horizontal and lined up with the hole in the motor mount.
3.
Remove the port plug using a 3/16" allen wrench.
4.
Fill the reservoir up to the bottom of the fill port.
5.
Replace the plug and run the cot up and down a few times.
www.stryker.com
Troubleshooting Guide
6500-009-005 REV A
1

Publicité

Table des Matières
loading

Table des Matières