Troubleshooting Guide
POWER PRO Electrical System Block Diagram
Lower and Retract (Load) Functions
Press (–) Button
Yes
Pressure Switch
Pressure Switch
Open B-Valve
Open?
Open?
No
Yes
No
Open B-Valve and Ramp in
Motor
Hold B-Valve open until
button released or pressure
switch closes
Run motor at max speed
No
Bottom Hall Effects
Sensor Detected?
Yes
Ramp out motor and stop,
Close B-Valve