The following is a list of safety precautions that must be observed when operating or servicing this unit. The precautions
are repeated throughout the manual, where applicable. Carefully read this list before using or servicing the unit.
WARNINGS
•
Do not modify the
Modifying the product can cause unpredictable operation resulting in injury to the patient or operator. Modifying the
product will also void its warranty.
•
The in-fastener shut-off module must be positioned properly before placing the ambulance cot into EMS service.
Failure to install the in-fastener shut-off module may cause injury to the patient or operator and/or damage to the
vehicle.
•
It is the responsibility of the cot operator to ensure the ambulance cot being used in the Stryker cot fastener system
meets the installation specifications listed on page 20. Injury may result if a non-compatible ambulance cot is used
in the Stryker cot fastener system.
•
Have the vehicle safety hook installed by a certified mechanic. Improper safety hook installation can result in injury
to the patient or operator and/or damage to the cot.
•
Failure to install the vehicle safety hook can result in injury to the patient or operator. Install and use the safety hook
as described in this manual. To avoid injury, verify the safety bar has engaged the safety hook before removing the
ambulance cot from the patient compartment.
•
Install the vehicle safety hook at least 1/8" (3 mm) from the edge of the rear ambulance door. After installation,
verify that the undercarriage locks in the load position without contacting the emergency vehicle bumper.
•
The ambulance cot must have at least 5/8" (1,6 mm) of clearance between the emergency vehicle bumper and the
ambulance cot to disengage the safety bar when unloading the product from the ambulance. Verify that the product
is locked in a rolling position before disengaging the safety bar from the safety hook (if equipped).
•
To avoid accidental release of the Pedi-Mate™, and possible injury to the infant, ensure that the restraint buckle is
located away from obstructions on the ambulance cot or accessories.
•
Never apply the optional wheel lock(s) while a patient is on the ambulance cot. Tipping could occur if the ambulance
cot is moved while wheel locks are applied, resulting in injury to the patient or operator and/or damage to the
product.
•
Never install or use wheel locks on an ambulance cot with excessively worn wheels. Installing or using wheel locks
on wheels with less than a 6" (16 cm) diameter could compromise the holding ability of the wheel lock, resulting in
injury to the patient or operator and/or damage to the product or other equipment.
•
To avoid risk of electric shock, never attempt to open the battery pack for any reason. If the battery pack case
is cracked or damaged, do not insert it into the charger. Return damaged battery packs to a service center for
recycling.
•
Do not remove the battery when the ambulance cot is activated.
•
Avoid contact with a wet battery or battery enclosure. Contact may cause injury to the patient or operator.
•
When the optional head-end storage flat is being used, ensure it does not interfere with the operation of the
retractable head section, safety bar and safety hook. Injury to the patient or operator could result.
•
Improper usage of the
ambulance cot only as described in this manual.
•
Entanglement in powered ambulance cot mechanisms can cause serious injury. Operate the ambulance cot only
when all persons are clear of the mechanisms.
•
Practice changing height positions and loading and unloading the ambulance cot until operation of the product is
fully understood. Improper use can cause injury.
•
Do not allow untrained assistants to assist in the operation of the ambulance cot. Untrained technicians/assistants
can cause injury to the patient or themselves.
•
Do not ride on the base of the
injury to the patient or operator.
•
Grasping the
moving parts.
www.stryker.com
Summary of Safety Precautions
ambulance cot or any components of the cot, including the hydraulic unit.
ambulance cot can cause injury to the patient or operator. Operate the
ambulance cot. Damage to the product could occur, resulting in
ambulance cot improperly can cause injury. Keep hands, fingers and feet away from
6500-009-005 REV A