Chapter 3: Setting Up
This chapter contains information about setting up NomoLine ISA CO2 before use.
System Setup
During setup, refer to the host device operator's manual or user's guide, in addition to the
following procedure.
1.
Remove the temporary plug from the NomoLine Capnography input connector.
2.
Securely mount the NomoLine ISA CO2.
Note: A bracket designed to mount NomoLine ISA CO2 is available, visit
www.masimo.com.
3.
Connect the gas sample exhaust port on the rear of NomoLine ISA CO2 to a
scavenging system if intended to be used in combination with N2O and/or
anesthetic agents.
4.
Connect the NomoLine ISA CO2 cable to the host device connection port.
5.
Connect a NomoLine sampling line to the NomoLine ISA CO2 input connector. See
Front View on page 15.
6.
Check that host device is powered up and correctly configured.
7.
Check that the LEGI indicator shows a steady green light, indicating that
NomoLine ISA CO2 is ready for use. See LEGI Indicator on page 19.
Note: Without a sampling line connected, the LEGI Indicator does not illuminate.
8.
Attach a NomoLine sampling line to the patient for monitoring. Refer to the
NomoLine sampling line Directions for Use.
9.
Following connection of the NomoLine sampling line, check that CO
appear on the host device screen. See Capnography Display on page 19.
www.masimo.com
values
2
17
Masimo
Chapter 4: Operation
The information in this chapter assumes that NomoLine ISA CO2 is set up and ready for use.
Do not operate NomoLine ISA CO2 without completely reading and understanding these
instructions.
LEGI Indicator
The Light Emitting Gas Inlet (LEGI) Indicator provides visual indications of capnography
status. The LEGI Indicator (3) is located around the capnography connector on the front of
the device.
The LEGI indicator illuminates in different colors depending on the state of the device as
described in the table:
LEGI Indicator
Status
Steady green light
Capnography monitoring in operation and OK
Blinking green light Zeroing in progress. See Zeroing on page 34.
Steady red light
Sensor error
Blinking red light
Check the sampling line (possible occlusion)
Note: Without a NomoLine sampling line connected, the LEGI Indicator does not illuminate.
With no sampling line connected, NomoLine ISA CO2 stays in a low-power, sleep mode. Once
a sampling line is connected, NomoLine ISA CO2 switches to measuring mode and starts
delivering gas data.
Capnography Display
Parameters and measurements display on the screen of the host device that NomoLine ISA
CO2 is connected to. Refer to the host device operator's manual or user's guide for complete
information on displayed parameters, settings, alarms and messages.